Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 21: CAL ADJ /5700 INSTRUMENT: Fluke 23: CAL ADJ /5700 INSTRUMENT: Fluke 73: CAL ADJ /5700 INSTRUMENT: Fluke 75: CAL ADJ /5700 INSTRUMENT: Fluke 77: CAL ADJ /5700 DATE: 28-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 54 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # Fluke 70 Series service manual (PN 731034 Jan. 1984). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 57_7xCAL.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R W 1.002 HEAD CALIBRATION ACCESS PROCEDURE 1.003 DISP Select the V DC function (auto range). 1.003 DISP 1.003 DISP Remove the four phillips-head screws from the bottom of 1.003 DISP the case. Turn the instrument face-up, and grasp the top 1.003 DISP cover at both sides. Pull the top cover from the UUT. 1.004 HEAD EQUIPMENT SETUP 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Allow the UUT to stabilize. 1.006 HEAD CALIBRATION PROCEDURE 1.007 PIC 57_7xCAL 1.008 5700 3V S 2W 1.009 DISP Adjust the Calibration Pot for a reading of 3.000V. 1.010 END